Add to Favorites
Share

Digital Portable Central Monitoring System Central Monitoring Device

  • Description
Product Detail

    PRODUCT DESCRIPTION

    Digital portable central monitoring system central monitoring device  

    Landcom digital portable central monitoring system central monitoring device model LD-8 is a full-featured system that provides comprehensive patient monitoring and review, it is vital in operating and emergency room, as well as intensive and critical care settings.  

     

    FEATURES

    1. 12.1' color TFT LED screen supports display max. 11 waveforms.  

    2. Personated voice reminding alarm; visual on-display and integrated alarm light.   

    3. Display mode: standard, large font, NIBP trend, trend table, Oxy CRG, full leads ECG. 

    4. Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection. 

    5. Drug calculation, titration table.

    6. Suitable for monitoring adult, pediatric and neonatal patient.

    7. Extended advance function: 2IBP, IOC, EtCO2 , Multi-gas/O2. 

    8. Supporting storage of 720 hours patient data, 1000NIBP records and 500 alarm event.

    9. Multi-language: Chinese, English, Spanish, Portuguese, French, German, Czech, Finland, Dutch, Italian, Porlish, Turkish.

    10. Anti-glare screen, rechargeable battery, 3/5-Lead ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature.

    11. 2IBP, Nellcor SpO2, Depth of anesthesia monitoring module, EtCO2 (Main-stream/Side-stream/Micro-stream CO2), multi-gas/O2, pitch tone.

    12. VGA, Thermal printer (50 mm).

    13. Pediatric/neonatal accessories: reusable NIBP cuffs & SpO2 probe; disposable NIBP cuff & SpO2 probe.

    14. Central Monitoring System.

    15. Touch Screen.

     

    PHYSICAL PROPERTIES

    Physical Specifications

    Dimension and weight

    Dimension

    340mm(W)*280mm(H)*130mm(D)

    Weight

    Standard configuration: 4 kg

    Operation environment

    Power requirements

    AC 90-240V, 50/60HZ, 60W

    Power

    90VA

    Temperature

    0-40?

    Humidity

    0-93% non-condensing

    Safety

    Meet the requirements of IEC60601 series

    Patient range

    Adult, neonate and pediatric patients

    Performance Specification

    Display

    12.1’ (800 x 600 pixel) color TFT LED screen, support external VGA, rolling and refreshing waveform display

    Display mode

    Standard, big font, NIBP trend, trend table, full-lead ECG, Oxy CRG.

    Interface

    Parameter cable interface

    AC power input socket

    Network interface

    Battery

    1 external rechargeable lead-acid battery, 2.0Ah 12V

    Alarm indicator

    Visual on-display and integrated alarm light

    Audible alarm and personated voice reminding alarm.

    Data storage

    720 hours monitoring data

    1000 NIBP records storage

    500 alarm events

    Printer

    Built-in 50mm thermal printer (option),

    3 traces, print speed 12.5mm/s & 25mm/s & 50mm/s

    Print width: 50mm

    Net working

    Support connection to central monitoring station through LAN (TCP/IP) or wireless connection

    ECG

    Leads type

    3/5 lead : I, II, III, aVR, aVL, aVF, V

    Sweep speed

    12.5mm/s, 25mm/s,50mm/s

    Gain selection

    ×0.25, ×0.5, ×1, ×2

    ECG waveform

    3 channels

    Operation mode

    Diagnostic, monitor, surgery

    Heart rate range

    15 - 350bpm

    Accuracy

    1bpm or ±1% whichever is greater

    Alarm range

    Audible and visual alarm, adjustable

    ST/ arrhythmia

    Real time S-T segment detection, pace detection and arrhythmia analysis

    Protection

    Against electrosurgical interference and defibrillation

    Respiration (RESP)

    Method

    Thoracic impedance

    Operation mode

    Auto/manual


    Measurement range

    Adult: 0-120BrPM; neonatal /pediatric: 0-150BrPM

    Apnea alarm

    Yes


    Alarm

    Yes, audible and visual alarm ,alarm events recallable

    NIBP

    Method

    Automatic oscillometric

    Operation mode

    Manual / Automatic / STAT

    Auto measure time

    Adjustable

    Measurement unit

    mmHg / kPa selectable

    Test interval

    1-480 minutes

    Measurement range

    Adult: 0-300 mmHg ;

    Child: 0-240 mmHg ;

    Infant: 0-150 mmHg

    Over-pressure protection

    Double safety protection

    Accuracy of blood pressure      measurement

    The mean error shall be less than ± 5mmHg

    The standard deviation shall be less than 8mmHG

    Resolution

    1 mmHg

    SPO2

    Measurement range

    0%-100%

    Resolution

    1%

    Accuracy

    ±2% (70-100%, adult/pediatric, non-motion )

    ±3% (70-100%, neonate, non-motion )

    ±3% (70-100%, adult/pediatric/neonate, motion )

    0-69% unspecified

    Alarm range

    1-100%


    Pulse rate

    Pulse range:

    20-300bpm

    Accuracy

    ±3bpm(non-motion)

    ±5bpm(motion )

    Temperature (TEMP)

    TEMP range

    0-50?


    0.1 ?

    Accuracy

    ±0.1?  (excluding the sensor)

    Channel

    Single channel

    IBP

    Measurement range

    -50-360mmHg

    Channel

    2 channels

    Pressure transducer

    Sensitive :5 uV/V mmHg

    Impedance range :300-3000 O

    Pressure names

    ART, PA, CVP, RAP, LAP, ICP, P1, P2

    Alarm range

    -10-300 mmHg

    PR from IBP

    Range & Accuracy:

    25-200bpm  ±1bpm or ±1bpm which is greater

    201-350bpm ±2%

    Resolution

    1bpm

    Depth of Anesthesia (BIS)

    EEG sensitivity

    +/- 475µV

    Noise

    < 2uVp-p (1-150 Hz)

    CMRR

    > 100 dB

    Input impedance

    > 50mohm

    Sample rate

    1024 samples/s, 16 bits

    Index and display update

    0-100, filter 1-47 Hz, 1 sec. display update

    EEG waveform

    ± 500 µV scaleable, filter 1–30 Hz, 3 sec. window length

    ESR

    0-100%

    EMG 0-100 arbitrary unit

    Curve and bar

    Index trend

    5 min and 60 min

    Digital output

    Wireless BT interface (RFCOMM) 10 meters range

    Alarms

    IoC high and low user selectable limits

    Artifact rejection

    Automatic: user can disable.

    Sensor impedance range

    0 - 14kohm

    EtCO2 (Main-stream / Side-stream )

    Measurement range

    0-150mmHg

    Accuracy

    ±2mmHg(0-40mmHg)

    ±5% of reading(41-76mmHg)

    ±10% of reading (77-99mmHg)

    Resolution

    1mmHg

    Sampling rate

    70m1/min 100m1/min

    Sampling rate accuracy

    0.15 or 15m1/min

    Start-up time

     

    30 Sec. the module enters the warming up status after the start up.1 minute later, it enters the ready-to-measure status.

    Respiration rate

    3-150 rpm

    Respiration rate accuracy

    ±2BrPM (0-70BrPM)

    ±5BrPM (>70BrPM)

    Response time

    When measured with a neonatal water trap and a 2.5 m-long neonatal sampling line:

    < 3.5 s @ 100 ml/min

    <4 s @ 70 ml/min

    When measured with an adult water trap and a 2.5 m-long adult sampling line:

    <5.5 s @ 100 ml/min

    <7 s @ 70 ml/min

    Delay time

     

    When measured with a neonatal water trap and a 2.5m-long neonatal sampling line:

    <3 s @ 100 ml/min

    <3.5 s @ 70 ml/min

    When measured with an adult water trap and a 2.5m-long adult sampling line:

    <5 s @ 100 ml/min

    <6.5 s @ 70 ml/min

     

    WARRANTY PERIOD

    Products’ warranty period is 1 year from the goods delivery. 

    CERTIFICATES

    Our certificates are as below

    89.jpg

    RELATED PRODUCTS

    related products.jpg

    PACKING

    56.jpg

    PAYMENT &SHIPMENT

    12.jpg


    OUR ADVANTAGES

    1.More than 20 years for manufacturing medical equipments.

    2.Tsinghua technology forges the brand of Landcom, and Tsinghua University, Xi’an Jiao Tong University and 

    Northwest University are the backbones of Landcom R&D center. 

    3.Landcom joined Chinese medical instrument industry association and Chinese automobile industry association,

     and became one of the directors. 

    4.Landcom passed ISO9001, ISO13485, CE and WMI code. 



    OUR SERVICES 

    Reliability and the tenderest service are provided to clients.

    1.Landcom supply value-added services to clients and to update the product software free for clients. 

    2.Landcom supply all-around response service & to reply clients’ demands in 24 hours.

    3.Landcom have more than 20 experienced engineers service & solve problems for clients. 

    4.According to the official order quantity and actual situation of importing country, Landcom can build an 

    after-sale service station in importing country to service clients. 


    OUR COMMONWEAL AFFAIRES

    1.Landcom donated 6 units FORD ambulance during the SARS crisis in 2003. 

    2.Landcom donated some computers to “xiwang” schools in 2003. 

    3.Lancom donated 2 units Mercedes Benz ambulances for Wenchuan earthquake disaster in 2008. 

    4.Landcom donated 2 units IVECO ambulances for Qinghai earthquake disaster in 2010. 

    23.jpg

    COMPANY INFORMATION 


    Office building

    01.jpg

    Warehouse

    02.jpg

    Clients’ visit 

    03.jpg

    Welcome to get superior and competitive price digital portable central monitoring system central monitoring device for sale with Landcom Medical, a professional distributor and wholesaler in China. As one of the leading manufacturers, we have established reliable cooperative relationships with many companies.